Routine Care
Angel
®
cPRP System
Operator’s Manual
6-7
DISINFECTION
Using a disinfecting towelette or a clean, low-linting cloth dipped in disinfectant solution, gently wipe
down all surfaces of the
Angel System
. Ensure all surfaces remain visibly wet for the contact time
recommended by the disinfectant manufacturer. If required by the disinfectant manufacturer, rinse per
instructions; otherwise, allow to air dry.
Recommended disinfectants with given compatibility with the
Angel System
include quaternary-based
compounds, OPA (ortho-phthalaldehyde), and chlorinated solutions (bleach at dilutions of 1:15 to
1:100) that are EPA-registered to achieve intermediate-level disinfection.
To prevent possible discoloration or damage to the protective panels on the
Angel System
, do not use
abrasive disinfecting products such as iodophors (iodine, Betadine®), benzol, solvents (xylene,
acetone, benzene), alcohols, or glutaraldehyde.
GENERAL
Routinely verify that all air intake areas of the
Angel System
are clear of debris and clean as required.
If the device is equipped with screens on the bottom of the unit, periodically inspect the screens for lint,
debris and/or damage. As necessary, remove the lint or debris using a vacuum to clean the screens
externally, taking care not to damage the screens.
Note:
Not all models are equipped with screens. If damage is observed, contact Arthrex, Inc. Customer
Care at + 1 800-391-8599 or [email protected].
Note:
Only the external portion of the screen is to be cleaned, as required, by the customer. The unit
should never be opened by the customer.
Non-Routine Care
In the Event of a Serious Fluid (Blood) Spill
Fluid (blood) spills that accumulate in the Centrifuge Well of the
Angel System
require different
procedures than spills isolated to the exterior of the machine. In the event of this type of fluid (blood)
spill, steps will need to be taken to disassemble and clean the Centrifuge Assembly and the Centrifuge
Well prior to returning the
Angel System
to Arthrex. Follow these steps carefully to properly clean these
kinds of spills and to ensure a biohazardous situation is avoided during the return.
Dissassembling & Cleaning the Centrifuge Parts & Centrifuge Well
Using a sponge or cloth rinsed in mild detergent, disinfectant or a 15:1 water to bleach solution, absorb
and discard any excess fluids that have accumulated in the Centrifuge Well.
Once the bulk of the fluids have been removed, follow the procedure below to remove, disassemble and
clean all parts of Centrifuge Assembly:
PRECAUTION
Be careful not to misplace the screws and springs when disassembling the Centrifuge
Adapter Plate.
Note the removal pattern illustrated in Figure 6-7 below.
Summary of Contents for Angel
Page 2: ...This page intentionally left blank ...
Page 8: ...Overview vi Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 14: ...Overview xii Angel cPRP System Operator s Manual ...
Page 17: ...Overview Angel cPRP System Operator s Manual xv This page intentionally left blank ...
Page 18: ......
Page 26: ...Overview 1 8 Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 66: ...Processing 3 28 Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 98: ...Routine Care 6 12 Angel cPRP System Operator s Manual This page intentionally left blank ...