Instructions for Use, AquaVENT
®
FD140, Rev. 12
79
Apnoea
20 s
P max
12 cm H
2
O
P min
2 cm H
2
O
F max
OFF
9.6 Measuring functions
Increment
Min.
value
Max. value
Accuracy
Flow
1 l/min
0 l/min
200 l/min
± 10%
Respiratory
rate
1 /min
0 /min
60 /min
± 2 /min
Patient
pressure
(actual value)
1 cmH
2
O
0 cmH
2
O
50 cmH
2
O
± 10%
Patient
pressure
(bar graph)
0.25 cmH
2
O
0 cmH
2
O
25 cmH
2
O
± 10%
9.7 Paramagnetic oxygen sensor
Accuracy:
± 2%
Calibration:
annual
Service life:
12 years
9.8 Display
Screen type:
LCD with multi-colour backlighting
Screen diagonal:
3.3 inch
Screen resolution:
160 x 104 pixels
9.9 Alarms
Type of alarm:
visual, audible
Volume:
77.5 dB(A) when set at 100%
Alarm pause:
120 s, yellow alarm OFF LED
Summary of Contents for AquaVENT FD140
Page 1: ...AquaVENT FD140 Instructions for use CRITICAL CARE...
Page 6: ...0 Contents 4 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Page 11: ...Instructions for Use AquaVENT FD140 Rev 12 9 2Safety In this chapter 2 1 Safety precautions 10...
Page 16: ...2 Safety 14 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Page 56: ...5 Operation 54 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Page 70: ...7 Cleaning 68 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Page 76: ...8 Maintenance 74 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Page 95: ...Instructions for Use AquaVENT FD140 Rev 12 93 This page is intentionally blank...