Testing Procedure:

1. 

Add whole blood to the fill line on the centrifuge cup or serum / plasma to fill line 

     on serum cuvette.
2. Press the [1] key on the Main Menu to initiate a measurement.
       • The table cover will open and the STAND-BY screen is displayed.

3. 

Press the [6] key to bring up and scroll through the species menu.

4. 

Press [ENTER] to select the desired species.

       • Unit will go back to the STANDBY screen
5. Press the [1] key for ID.
6. Enter the test number (if desired) otherwise press [ENTER].
7. ID option will appear - enter patient number or name, press [ENTER].
       • The [+/.] key is the space key and can be used to move the cursor to the next number/   
           letter key.
8. When using serum/plasma, place the serum cup in the serum holder.
       • Skip to step 12.
9. Remove centrifuge cover by pulling backwards and upwards.
10. Place the centrifuge cup with whole blood in position on the indented area of the silver 
      tabs.
11. Install centrifuge cover.
12. Place the pipette tip in the Tip Hole.
13. Place the reagent strips on the reagent table.
       • Reagent pads on single strips should be facing up and aligned towards the rear of the  
          instrument.
       • Reagent pads on multi strip should be facing up and aligned towards left side of the    
          instrument.
14. Press [START] key to begin measurement.
       • When the measurement is completed, remove used strips/centrifuge cup/cuvette and  
           discard them.
       • Instrument is ready to run the next sample.

How to run a sample:

E02

(T): Trouble with the table Cover.

(M): Trouble with the maintenance Cover. 

•  The measurement table was open during measurement or 

warm up

•  The maintenance cover was opened during measurement 

or warm-up

•  The table cover is not properly closed (foreign material may 

be preventing closing)

•  The measurement was started without closing the 

            maintenance cover

•  Close the table cover correctly (remove obstructions)

•  Set the maintenance cover in the correct position

E05

(1~9): Channels with abnormalities

•  A check shows the optical system status has changed since 

the last calibration

•  The white and black plate or Optical window (window plate) 

is dirty

•  Clean the white and black plate or Optical window (window 

plate) - see sections 4-2-1 and 4-3-2 in the owner’s manual

E11

(7~9): Single Type                (M): Multi type

Channel with no Reagent strip in either case. Display is made only in 

calibration. Usually nothing is displayed during measurement. 

•  Reagent strip is not set

•  The bar code was not correctly read due to displacement or 

bending of the Reagent strip

•  The Reagent strip is not set in the channel necessary for 

calibration

•  Set Reagent strips correctly

•  When calibrating, swipe all necessary numbers on card

E13

(7~9): Single Type                (M): Multi type

Channel with a different Reagent strip in either case.               

•  The bar code was not correctly read due to displacement or 

bending

•  The wrong reagent strips was set for calibration

•  Calibration was started for an item before it was completed 

for another item

•  Set the Reagent strip correctly

•  Return to the main menu because changing the item          

mid-calibration is not allowed when doing the calibration

E14

(7~9): Single Type                (M): Multi type

Channel has used Reagent strip in either case.            

•  Reagent strip is already used

•  The Reagent strip pad is colored because of an old Reagent 

strip of inappropriate preservation

•  The Reagent pad is dirty

•  The bar code was not correctly read due to displacement of 

bending of the Reagent strip

•  Use a new Reagent strip

•  Set the Reagent strip correctly

E17

•  A sample vessel is misplaced

• 

There is no sample or an insufficient amount of sample

•  A centrifuge tube was used in calibration by calibrator kit

•  Re-set the sample correctly

• 

Confirm the amount of the sample

•  Be sure to use cuvettes in calibration by calibration kit

E32

•  Reagent strip has expired.

•  Clock is not correctly adjusted

•  Clock is not correct because the power was not connected 

for a long time

•  When using a Reagent strip from new lot, the corresponding 

magnetic card to the lot was not inserted

•  Use a new Reagent strip

•  Adjust the clock

•  Insert a new magnetic card

E38

•  The tip waste receptacle is not correctly set

•  Set the tip waste receptacle correctly

E40

(1~9): Channels with abnormalities

•  Failed to draw the sample and drop it to the Reagent 

correctly due to lack of sample amount or absorbing fibrin

•  Sample adhered around the dropping hole

• 

Check the sample amount and remove fibrin if necessary

•  Wipe off the sample adhered around the dropping hole

•  Measure with a different tip

T25

(1-5, 7-9): Channels with abnormalities

•  White plate or black plate is dirty

•  Clean the white plate or black plate. (Note: the case in 

which trouble is detected when the power is turned on, it 

is necessary to turn on the power again after maintenance 

since measurement cannot be started)

T03

(1-5, 7-9): Channels with abnormalities

•  The pressure does not increase since the rubber plate is 

dirty or deformed

•  The tube or pipe is clogged

•  The tube connecting the nozzle with the pump is 

disconnected

•  The tube connecting the pump with the pressure detecting 

sensor is disconnected

•  Clean the rubber plate

•  Check the tube

•  If the same trouble occurs, contact your distributor (Note: 

in cases where the abnormality is detected when the power 

switch is turned on, it is necessary to turn on the power 

again after maintenance because the measurement could 

not be started)

E02 Cover Open

  OK 

(ENTER)

E05 Optical Error

(9     6     2)  

OK (ENTER)

E11 No Strips

(                M) 

OK (ENTER)

E13 Wrong Strips

(  87         ) 

OK (ENTER)

E14 Used Strips

(9                     ) 

OK (ENTER)

E17 No Sample

(CNTRFG) 

OK (ENTER)

E32 Invalid Strips

(    7        M) 

OK (ENTER)

E38 Tip Case Not Set

  OK 

(ENTER)

E40 Sample Drop

(   7   4   1) 

OK (ENTER)

T25 Barcode Sensor

(8   5   3)  

OK (ENTER)

T03 Tube Pressure

  OK 

(ENTER)

Troubleshooting

Description and Error Message

Conditions and Causes

Corrective Action

Centrifuge Cover

Place the Pipette Tip in Holder

Place Test Strip on Reagent Table

fill line

max 500 �L

fill line 250 �L

fill line

min 100 �L

Contact scil animal care toll free at 877.724.5838

Contact scil animal care toll free at 877.724.5838

SPOTCHEM

 EZ Quick Instruction Guide

SPOTCHEM

 EZ Quick Instruction Guide

scil animal care company     •     151 N. Greenleaf St. Gurnee, IL 60031   •     P: 877-724-5838  •     www.scilvet.com

scil animal care company     •     151 N. Greenleaf St. Gurnee, IL 60031   •     P: 877-724-5838  •     www.scilvet.com

Whole Blood – uses centrifuge cup

1.Using a lithium heparinized collection tube (recommended method)

       • Aspirate sample into syringe (no anticoagulant) 

       • Remove needle from syringe

       • Dispense sample into lithium heparin collection tube and mix sample

       • Transfer sample into centrifuge cup

2.  Direct transfer to centrifuge cup 

       • Aspirate sample into syringe (no anticoagulant) 

       • Remove needle from syringe

       • Dispense sample directly into centrifuge cup 

       • Re-install orange cap on centrifuge cup

       • Mix sample in centrifuge cup;

  verify  sample  is  completely  heparinized      

          before proceeding (minimum of 2 minutes)

Plasma/Serum – uses cuvette 

       • Prepare plasma/serum sample 

       • Transfer sample into cuvette tube

Sample Preparation:

Centrifuge cup Placement

Cuvelte

Centrifuge Cup