Chapter 3 Supplementary Operations
AE-4022
3-6
3.3
MEMORY
(reprinting and resending measurement results)
Use the MEMORY menu to reprint or resend measurement data (of up to 520 samples) stored in the instrument’s
memory. The stored measurement results are classified by measurement mode and by measurement result.
To reprint or resend the data, specify the “measurement period”, “measurement mode”, “sample”, and “measurement
result type” so that the desired measurement results will be selected for output.
The reprint/resend functions are available even if the settings for built-in printer use and external equipment use are
set to “OFF”.
1
Call up the menu screen
a
With the [Standby screen] displayed, press
.
• The [Menu screen] appears and
flashes.
2
Choose MEMORY (to reprint or resend the measurement results)
a
Press
or
several times until
flashes.
b
Press
.
• The [Output method selection screen] appears.
3
Select the desired output method
a
Press
or
several times until the desired output
method flashes.
•
: Resend
: Reprint
b
Press
.
• The output method is confirmed and the [Start date of
measurement period input screen] appears.
REFERENCE:
If you press
during operation or input, the current setting is canceled and the display returns to the [Menu
screen].