1.1 Outline of the AE-4022
AE-4022
1-5
1.1.3
Specifications
Sample
Urine
Test Strip / Reagent pack
AUTION Sticks 10EA
Measurement item
GLU, KET, BIL, NIT, PRO, URO, pH, BLD, LEU, Specific Gravity, color tone
Measurement range
Test strip: Rank table (See “6.1. Rank tables” on page 6-2)
Color tone: Color tone classification chart (See “Technology and Scientific Principles”
on page vi)
Measurement method
Test strip method
Dual-wavelength reflection photometric method
(single wavelength for BLD)
Measurement wavelength
4-wavelength LED (430, 565, 635, and 760 nm)
Reaction time
60 seconds
Processing speed
514 samples per hour
(maximum processing mode: 7-second interval)
Display
Custom LC display (icons are used)
Built-in printer
32-digit thermal printer (58 mm width)
External output
Conforms to RS-232C standard (serial)
Option: Ethernet
Memory capacity
520 measurement results
Operating environment
Temperature: 50 to 86°F (10 to 30°C); Humidity: 20 to 80% R.H.
(non-condensing)
Measurement environment
Temperature: 50 to 86°F (10 to 30°C); Humidity: 30 to 60% R.H.
(non-condensing)
*Temperature correction function used
Storage environment
Temperature: 33.8 to 86°F (1 to 30°C); Humidity: 20 to 80% R.H.
(non-condensing)
Environment during
transport
Temperature: 14 to 160°F (-10 to 60°C); Humidity: 20 to 80% R.H.
(non-condensing)
Site location
Indoor use only
Altitude
Up to 2000 m/ 6560 feet
Pollution degree
2
DC power requirements
(To instrument)
12 Vdc 3 A
AC power requirements
(To AC adapter)
100-240 Vac 50-60 HZ ,1200 mA
Voltage fluctuation allowance is ±10%
Power consumption
Max. 45 VA
Overvoltage category
II
Dimensions
210 (width) × 328 (depth) × 164 (height) mm
Weight
Approx. 3.6 kg
Expected life
5 years from first use (installation) of the instrument (according to company data).