Lifting poles CM-ACC09
The lifting pole is used to assist a patient in
moving or turning whilst lying on the bed.
Model CM-ACC09 is fixed in one position
above the center of the bed.
The lifting pole is supplied with an adjustable
strap and handle.
WARNING
Do not overload the lifting pole. The maximum weight it can
safely support is 165lb.(75kg).
Do not hang other objects on the lifting pole, strap or handle.
The bed should not be occupied while the lifting pole is being
installed.
Do not use the lifting pole to push or pull the bed.
The operational life of the strap and handle is two years when used and
maintained in accordance with the manufacturer’s instructions. After this
time, the complete unit should be replaced.
Inspect the strap and handle regularly. If any sign of wear or damage is
found, remove it from use immediately and replace the complete unit.
Only use the recommended strap and handle assembly - Avitana type
04120.
Do not adjust the strap while a patient is holding the handle as this may
damage the mechanism. After adjusting the strap, make sure it locks with
an audible “click”.
Do not twist the strap through more than 180
°
as this may cause
premature wear.
The lifting pole must be correctly positioned so that the
handle hangs over the middle of the bed. Do not allow the
lifting strap and handle to hang outside the perimeter of the
bed.
Ensure that the lifting pole and handle are positioned to suit
the size and medical needs of the patient.