11.
APPENDIX B – J405 FUNCTIONAL TEST KIT
If your application requires frequent verification of instrument functionality, this test will benefit
you. If the test results fall within the expected range, you may assume the instrument is
functioning properly.
THIS IS A FIELD CHECK OF THE FUNCTIONALITY OF THE INSTRUMENT.
THIS TEST DOES NOT CALIBRATE THE INSTRUMENT.
Testing with the Functional Test Kit (FTK) should always be done with the instrument set to
display in micrograms per cubic meter (µg/m
3
). If your instrument is displaying results in
milligrams per cubic meter (mg/m
3
), switch to µg/m
3
before doing any testing with the FTK. See
information on changing the display units.
NOTE: Perform the functional test ONLY after a sensor regeneration.
The J405 Functional Test Kit contains all accessories necessary to perform the functional test.
See the complete list on page
and verify that all the parts to the kit are present.
CAUTION
:
The vial and thermometer contain liquid mercury and are possible sources of mercury
contamination. Follow the instructions for handling or transferring the mercury into the
Functional Test Kit Vessel carefully.
For safety information, see the supplier’s Safety Data Sheets (SDS) or call AZI Customer
Service at 1-800-528-7411 or 1-602-470-1414 for assistance in obtaining the SDS.
Preparation
•
Carefully unpack and inspect the parts of the kit.
•
ENSURE that the mercury shipping container and mercury filled thermometer are not broken.
•
In a ventilated area, preferably under a fume hood, remove the mercury vial from its shipping
container.
•
Place the functional test kit vessel and the mercury vial close to each other and open the
mercury vial.