Internal Filters
Part Number
1. Scrubber Filter (1 elbow, 1 straight barb)
Z2600 3930
2. Scrubber Filter (2 elbow barbs)
600-0262
3. AG Filter
600-0293
Internal Filters
•
Replace the internal filters after six (6) months of use, or as needed.
The AG filter should be replaced every six (6) months or every 500 samples,
whichever comes first, based on your instrument usage schedule.
•
Press the
I/O
button to turn the power OFF and unplug the power cord.
•
Remove the two (2) side screws from the intake end of the instrument and open the case.
•
Carefully disconnect the Tygon
®
tubing from both ends of the filters and discard the old
filters.
•
Replace the old filters with new filters, oriented as shown in the diagram above.
Push the Tygon
®
tubing as far as it will go onto the filter fittings.
Note that the marked flow direction on the AG filter goes toward the sensor, or to the
left in the diagram above.
•
Push the filters into the mounting clips.
•
Remove any crimps or twists in the tubing and ensure that tubing connections are secure. If
the tubing is loose, readings may not be accurate. Replace any tubing that has deteriorated
due to heat and/or age.
•
Close the case and replace the screws.
•
If the AG filter is replaced, the new filter must be acclimated to the system by performing a
sensor regeneration, followed by twenty (20) manual samples using the sample button. See
page
for sensor regeneration instructions.
•
Dispose of all filters in accordance with state and federal environmental regulations.
#1
#2
#3