22 Copper tests
296
ARGUS 152
ARGUS first measures the resistance.
ARGUS determines the capacitance
when it detects an open line (infinite
resistance).
ARGUS displays the resistance. The
capacitance is not displayed because a
loop is used in this example. ARGUS also
determines the approximate line length,
see equivalent circuit diagram page 294.
Resistance test: 20
to 100 k
Accuracy: 20
≤
R
≤
100
: ±10 %
R > 100
: ±2 %
ARGUS displays the capacitance. The
resistance is outside the possible range
(> 100 k
).
Capacitance test: 1 nF to 1 µF
Test accuracy: ±5 %
Browse through the wire
types.
<Wire
types>
Switches to the wire types.
<New>
Repeats measurement.
Switches to status display.
Browse through the wire types.
<Wire
types>
Switches to the wire types.
<New>
Repeats measurement.
Switches to status display.
Loop:
Repeat test
Open line:
Repeat test