in#
Config.
in#
Config.
in#
Config.
0
Vab,Vba
8
Vcn, Vcs
27
Vb
1
Vbc, Vcb
9
N/A
26
In
2
Vca, Vac
10
N/A
3
N/A
11
N/A
4
N/A
12
Ia, In
5
N/A
13
Ib
6
Van, Vb, Vas
14
Ic
7
Vbn, Vbs
m#:
Magnitude correction 0 to
±
999999
p#:
Phase angle correction 0 to
±
359.9
s/r#:
0 to enter scale value; 1 to return scale value.
A.3.44
Store Setup (SST)
Command:
#SST
Stores the present instrument setup in Setup Location #.
# = 1 through 6
A.3.45
Setup, Recall (SUP)
Command:
#SUP
Recalls the instrument setup# as stored with Store Setup (SST).
# = 1 through 6
A.3.46
Timer, Arm (TAM)
Command:
TAM
Arms the Timer. For proper operation, the Timer must be armed prior
to a valid trigger condition. If there is a non-zero number on the Timer
display, the TAM command will set it to zero.
A.3.47
Time, Return (TDR)
Command:
TDR
Returns the present time in the format: hh:mm:ss dd mmm yyyy
A.3.48
Time, Set (TDS)
Command:
yyyy:mm:dd:hh:mm:ssTDS
121