Options List
C.9
Option 10: Inlet Power Supply Description
C.9.1
110 to 350 Vdc, 85 to 250 Vac, 47 to 440 Hz Terminal Power Strip, SWC
Option 10 replaces the standard IEC-320 power input module with a three-position, screw-type
terminal block, including Surge Withstand Capability (SWC). This feature is intended for use
in installations where it is necessary or desirable to have the instrument power hard-wired. See
Figure 3.3.
Figure C.9: Option 10 Power Supply Inlet
Connector
C.9.2
Specifications
Input Power:
AC Voltage Range:
85 to 250 Vac.
Frequency Range:
47 to 440 Hz.
DC Voltage Range:
110 to 350 Vdc.
Input Power:
<
20 Watts
Terminal Block:
Terminal Assignment:*
Ground, (
−
), (+), left to right, viewed from rear.
Block Size:
15 mm W x 18 mm H x 30 mm D. (5/8” x 0.75” x 1 3/16”).
Approvals:
U. L. recognized; C.S.A. approved.
*For AC operation, input line may be connected between (+) and (
−
), without
regard to polarity; however proper grounding should always be employed.
Fuse:
Type:
Bussman GDC-1A.
Current Rating:
1 Ampere, fast-acting.
Voltage Rating:
250 Volts.
Size:
5 mm x 20 mm.
Surge Withstand Protection (SWC)
Provides input Surge Withstand Capability (SWC) in compliance with both ANSI C37.90 and IEC
801-4.
Connections
All input power line connections to the rear-panel terminal strip should be made using appropriate
power cables which have the insulation removed about 1/4” from the end or as required for tinning.
If a DC source is used, connect the positive lead to the positive (+) terminal, connect the negative
lead to the negative (
−
) terminal and a safety ground lead to the “G” terminal when viewing the
instrument from the rear (see Figure 3.3).
Summary of Contents for 1084A
Page 4: ...iv ...
Page 6: ...vi ...
Page 18: ...xviii LIST OF FIGURES ...
Page 24: ...4 Unpacking the Clock ...
Page 36: ...16 Connecting Inlet Power Input and Output Signals ...
Page 44: ...24 GPS Antenna and Cable Information ...
Page 114: ...94 Startup and Basic Operation ...
Page 123: ...B 4 Physical Dimensions 103 Figure B 1 Suggested Mounting of the AS0094500 Surge Arrester ...
Page 124: ...104 Using Surge Arresters ...
Page 134: ...114 Options List C 6 2 Option 06 Firmware Setup Figure C 3 Option 06 Firmware Setup ...
Page 142: ...122 Options List Figure C 5 Option 06 Output Jumper Settings ...
Page 158: ...138 Options List Figure C 13 Option 17 Board Layout and Jumper Locations ...
Page 163: ...C 13 Option 20A Four Fiber Optic Outputs 143 Figure C 14 Option 20A Jumper Locations ...
Page 170: ...150 Options List Figure C 16 Option 23 Internal Jumper Setup ...
Page 172: ...152 Options List Figure C 17 Option 27 Jumper Locations ...
Page 190: ...170 Options List Figure C 20 Option 29 Connector Signal Locations ...
Page 246: ...226 Options List ...