![AR 10051217 Operating And Service Manual Download Page 65](http://html1.mh-extra.com/html/ar/10051217/10051217_operating-and-service-manual_2962298065.webp)
Model 2000S1G2z8
Rev A
41
2.13.5
Power Indicator Frequency
This command sets the current frequency used to correct the forward and reverse power indicators shown on
the display. The correction factors used by this feature are set from the factory and stored in non-volatile
memory.
Syntax:
CORR:POW:FREQ <freq><unit>
<freq>:
The current test frequency (0 to 999.999, limited to 3 digits of precision)
<unit>:
The frequency unit (KHZ, MHZ, GHZ)
Query:
CORR:POW:FREQ?
Example 1:
To set the frequency to 15.5MHz, send the following command:
CORR:FREQ 15.5MHZ<LF>
Example 2:
To query the frequency, send the following command:
CORR: POW:FREQ?
Response:
15.5MHZ<LF>
2.13.6
VSWR Limit Selection
This command sets the value of the output VSWR limit. If the limit is exceeded, a fault will be triggered, and
the RF output will be disabled.
Syntax:
VSWR:LIM <
selection
>
<selection>:
0 = None (Default)
1 = 6:1
2 = 3:1
3 = 2:1
Query:
VSWR:LIM?
Example 1:
To set the VSWR limit to 2:1, send the following command:
VSWR:LIM 3<LF>
Example 2:
To disable the VSWR limit, send the following command:
VSWR:LIM 0<LF>
Example 3:
To query the VSWR limit, send the following command:
VSWR:LIM?
Response:
0<LF>
Summary of Contents for 10051217
Page 2: ......
Page 4: ......
Page 6: ......
Page 18: ...Model 2000S1G2z8 iv...
Page 22: ...Model 2000S1G2z8 4 Rev A...
Page 85: ...Model 2000S1G2z8 Rev A 61 Figure 4 2 Typical Response at 0 dBm Input...
Page 86: ...Model 2000S1G2z8 62 Rev A...
Page 88: ...Model 2000S1G2z8 64 Rev A...
Page 90: ......