7. Calculate the temperature coefficient of sample solution according to the
formula shown above.
8. Enter the calculated temperature coefficient into the meter.
13.2 Appendix 2: How to calculate the TDS conversion
factor
To determine the TDS conversion factor, use the following formula:
Factor=
Where:
Actual TDS: value of a precisely weighed quantity of anhydrous NaCl or KCl
dissolved in high purity water.
Actual Conductivity: the meter measured conductivity value.
For example:
Dissolve 64 grams of potassium chloride reagent in 1L distilled water. If its
conductivity value is 100mS/cm, then TDS conversion factor is 0.64.
Actual TDS
Actual Conductivity @ 25ºC