EXIT THE CALIBRATION:
During the calibration process, if you want to exit calibration, press MEAS
key, the meter will return to measurement mode immediately.
10.1 Temperature Calibration
When automatic temperature compensation is enabled, the meter will need to
be calibrated if the temperature reading displayed differs from that of an
accurate thermometer.
10.1.1. Press and hold the MODE key for 3 seconds to enter temperature
calibration mode, the display shows current temperature reading.
10.1.2. Press ▲ or ▼ key to set the temperature value.
10.1.3. Press ENTER key to confirm. Calibration is completed.