7.3 Multi-Point Calibration:
7.3.1
Make sure that you have selected 2 to 5 points calibration in setup
menu.
7.3.2
Repeat steps 1.2 to 1.5 above; when the first calibration point is
completed, the display will show “CAL2”. The meter prompts you to continue
with the second calibration point.
7.3.3 Rinse the conductivity electrode with distilled water. Immerse the
electrode in the next calibration solution, the meter automatically shows
the currently detected calibration standard (e.g., 12.88mS/cm).
7.3.4
Press ENTER key to confirm, “Calibration” indicator begins
flashing.
7.3.5
Wait for the measured value to stabilise, the display shows “CAL3”.
The meter prompts you to continue with third point calibration.
7.3.6
Repeat steps 2.3 to 2.4 above until the display shows “END”. Once
calibration is complete, the meter returns to measurement mode
automatically.
Enter
Summary of Contents for WSL50 Pro
Page 1: ...AquaSafe WSL50 Pro Water Safety Laboratory Instruction Manual ...
Page 4: ...Section 2 Kit Contents c c c c b a e f d ...
Page 7: ...Section 3 ...
Page 8: ...AquaSafe Micro biological Incubator Instruction Manual ...
Page 10: ...6 0 Guarantee and Assistance 32 ...
Page 12: ......
Page 16: ......
Page 21: ......
Page 33: ...Section 4 ...
Page 34: ...AquaSafe Membrane Filtration Instruction Manual ...
Page 50: ...Section 5 HydroTest Photometer HT1000 ...
Page 86: ......
Page 136: ......
Page 142: ...HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Page 193: ...NOTES ...