5.3 Multi-Point Calibration:
5.3.1
Make sure that you have selected between 2 and to 5 point
calibration in the setup menu.
5.3.2 Repeat steps 5.2.2 to 5.2.5 above. When the first calibration point is
completed, the display will
show “CAL2”. The meter prompts to
continue with second calibration point.
5.3.3
Rinse the pH electrode with distilled water. Dip the electrode into
the next buffer solution (e.g., pH4.01). The meter automatically senses the
current calibration solution
and begins calibrating. “Calibration” indicator
begins flashing.
5.3.4
Wait for the measured value to stabilise; the display shows “CAL3”. The
meter prompts to continue with third calibration point.
5.3.5
Repeat the steps from 5.3.3 until the display s
hows “END”. Once
calibration is complete, the meter returns to measurement mode
automatically.