The maximum size of the <values> field is 75 characters when following a concurrent
measurement request, or 35 characters when following a non-concurrent measurement
request.
6.5.16. Continuous Measurements
The LeveLine
®
will respond to Continuous Measurements commands, but will not return
any measurements.
Command
Response
<a>R<set>!
<a><CR><LF>
•
<set> is a digit (“1”-“9”) indicating which set of measurements is being requested.
6.5.17. Continuous Measurements and Request CRC
The LeveLine
®
will respond to Continuous Measurements with CRC commands, but will not
return any measurements.
Command
Response
<a>RC<set>!
<a><crc><CR><LF>
•
<set> is a digit (“1”-“9”) indicating which set of measurements is being requested.
•
<crc> is a 3 byte string containing the encoded CRC of the response packet.
6.6. SDI-12 Device Address
The LeveLine
®
is assigned an SDI-12 device address, which is a single character from the
set “0”-“9”, “A”-“Z” & “a”-“z”. A LeveLine
®
will only respond to commands beginning with its
assigned device address or with the wild card address “?”.
The LeveLine's device address has a factory default value of “0”, and can be set to a
different value by a Change Address command.
6.7. SDI-12 Unit Identification
In its response to a Send Identification command, the LeveLine
®
will report compatibility
with SDI-12 version 1.3, the vendor ID “AQUAREAD”, the device model ID “LVLINE”, and
the LeveLine's firmware revision number and serial number.