8.4. Clearing the Memory
The memory within the Aquameter
®
is capable of storing over 1000 full sets of readings.
To clear the entire memory, switch the Meter off, hold down the
M+
key, then switch the
Meter back on. A screen will be displayed asking you to confirm your request. Press OK to
clear the memory or ESC to cancel and return to normal operation.
8.5. Automatic Data Logging
If you want to save readings on a regular basis, in order, say, to check water quality at a
certain location over a period of time, you can set the Meter to record readings
automatically.
Readings can be logged for short periods with the Meter permanently displaying readings,
or for much longer periods in a Low Power Mode, where the Meter switches itself off
between readings in order to extent the battery life.
Please note:
Low Power Logging Mode is only available on Meters running version 4.54
software and above.
To activate Automatic Logging, press the
MENU
key. The Main Menu screen will be
displayed. Please note, the first item on the menu, 'Clean Probe', will only be active if an
Aquaprobe
®
AP-7000 (which has an automatic cleaning system) is connected.
Clean Probe
Auto Data Logging
Calibration
Setup & Install
Select
Auto Data Logging
by pressing the down arrow key then the right arrow key or the
OK
key. The Auto Data Logging screen will be displayed.
Auto Data Logging
Interval:10 Mins
Status:OFF
Using the arrow keys to navigate, set the desired logging interval anywhere between 1 and
90 minutes.
To select permanent display logging mode, set the Status to
ON
. To select Low Power
logging mode, set the Status to
LOW POWER
.
To activate the selected logging mode, press the
OK
key then revert back to the normal
operation screen from the Main Menu by pressing the left arrow key.
To indicate that Auto Data Logging is switched on, an asterisk (*) character will flash on
and off just below the battery symbol on all the main reading screens.
If permanent display logging mode was selected (Status set to
ON
), the Meter will record a
full set of data automatically at the set rate until either the memory is full or the batteries go
flat.