1. Introduction
This manual covers the setup, operation, calibration and maintenance of the Aquaprobe
®
AP-5000 V4.07+, Aquameter
®
V6.20+ Meter, AquaLink V5.00+ PC software and associated
Aquaprobe
®
accessories. If your Aquaprobe
®
or Aquameter
®
are running earlier firmware,
the functionality may differ from that shown in this manual. In this case, contact Aquaread
for an earlier version of this manual or return your equipment for firmware upgrade.
2. What’s in the Box?
The Aquameter
®
is supplied with the following:
The Aquameter
®
unit.
Quick release lanyard.
Set of 5 AA Alkaline batteries.
USB Cable for downloading logged data to a PC.
Cross-head screwdriver for fitting the batteries and Probe maintenance.
Getting started card for quick reference.
The AP-5000 is supplied with the following:
Protective Sleeve End Cap (pre fitted).
600mL of RapidCal Solution.
Two part calibration / rinse cup.
One mounting nut (pre-fitted).
To complete your system, you will also need an AP-5000 Extension Cable, which should be
purchased separately.
2.1. The Aquameter
®
and the Environment
The Aquameter
®
is designed to be used outdoors and is rated to IP67, that is to say it is
waterproof but it
is not
designed for submersion. In order to prevent accidental dunking or
loss, a lanyard is supplied.
Please note that the socket on the Aquameter
®
is only waterproof when the
associated plug is fitted. Without the plug fitted, water can enter the socket. Damage
caused by water ingress through the socket is not covered by your warranty.
You may notice a small hole on the rear of the unit near the top. This is a waterproof vent
for the internal barometric sensor.
Do not poke anything in this hole!
Doing so will cause
major damage to the vent’s waterproof membrane and invalidate your warranty.
2.2. The AP-5000 and the Environment
The AP-5000 is designed to be fully submerged in water and is rated to IP68, that is to say,
it is rated for continual immersion to a depth of 30 meters, and short term immersion (less
than 12 hours) to 100 meters. During deployment, the AP-5000 can be suspended by the
Extension Cable, but in order to provide a fail-safe for the plug and socket, always use a
mounting bracket and Ty-wrap the Extension Cable to the bracket just above the plug.
2.3. Important Notes Regarding Galvanic Corrosion
Galvanic corrosion, sometimes also known as bimetallic corrosion, is an electrochemical
process in which one metal corrodes preferentially when it is in contact with a dissimilar
metal in the presence of an electrolyte (such as water). A similar galvanic reaction is
exploited in batteries to generate an electrical voltage.
© 2019 Aquaread
®
Ltd.
www.aquaread.com
Page 8 of 138