www.aqua-free.com
Instructions for use
1 Product
1.1 Product description
6 Accesories for Germlyser
®
IL 1500
- Ø 8mm Connection hoses (red/blue)
• Quick release adapter
• Isolation valve
• Locking ring
Common accessories are to be found in the current accessories
catalog or at www.aqua-free.com.
7 Maintenance and cleaning
The Germlyser
®
IL 1500 requires no maintenance within the specified operating time. The outside of the
filter can be treated by common surface disinfectants. Replace the connecting hoses, if necessary, for
example in case of leaks. For longer periods of non-use, it is recommended to close the water supply.
In the case of a deviation from the specifications or a damage of the packaging, the product should be
returned to Aqua free, along with a Defect Report. The Defect Report form is available to download from
the service page of the Aqua
free
website (www.aqua-free.com).
Please note
: Aqua
free
operates its own replacement service, which can be used by facilities with 25, or
more, filter points of use.
GEBR-ANW-IL1500-GB-0
8
/20
21
-V00
4
9 Returning
The Germlyser
®
reusable filters must be sent back to Aqua
free
within the agreed time for reprocessing
(form downloads at the service page under www.aqua-free.com).
8 Storage
The Germlyser
®
filter must be properly stored in a dry and hygienic environment at a temperature between
10 and 50°C.
Germlyser
®
IL 1500
Table of contents
1 Product
2 Certificates / Declaration of conformity
3 Labelling
4 Preparation of commissioning and use of product
5 Disassembly of used filter / filter change
6 Accessories
7 Maintenance / cleaning
8 Storage
9 Return
Headquarters:
Aqua free GmbH
Winsbergring 31
22525 Hamburg
Germany
Phone:
+49 (0)40-46 89 99 90
Fax:
+49 (0)40-46 89 99 99
www.aqua-free.com
Aqua free Solutions Ltd.
The Media Centre
7 Northumberland Street
Huddersfield, HD1 1RL, Great Britain
E-Mail:
Phone:
+44 (0)1484 483045
The Germlyser
®
IL1500 filter is a -marked, inline, sterile filters for use in cold water inlets of medical
equipment to prevent bacterial nosocomial infectious diseases caused by
Pseudomonas aeruginosa,
Legionella pneumophila, Stenotrophomonas maltophilia
or
Staphylococcus aureus
among immunosuppressed
and immunocompromised patients.
To achieve this intended use, a specific combination of retention of these
bacteria in the filter membrane structure and specific measurements to prevent retrogarde contamination with
these bacteria on the filter outlet, are necessary.
1.2 Technical data
Size:
L 122 mm x Ø 43 mm
Retention
*
:
7 log steps
Brev. dim.
Pore size:
0.2 µm
Operating time*
*
:
Depending on the application
<10 ppm
Chlorine resistance*
*
*:
Minimum flow rate:
6 l/min at 5 bar, at 25°C
Maximum operating pressure: 5 bar
Connection:
Quick-release coupling
*
*
The operating time depends on application conditions like water quantity and water quality.
*
** Continuous dosage of ≤ 10 ppm over operating time; short-term (1h) high dosage (400,000 ppm)
for chemical disinfection.
* Measured in accordance with the international ASTM F838-15a standard - Standard Test Method for
Determining Bacterial Retention of Membrane Filters Utilized for Liquid Filtration