CFS300 SERIES OPERATION MANUAL
SECTION 5: UNPACKING & INSTALLATION
Entire Contents Copyright
2016 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CFS300 Series Power Source Operation Manual
Page 54 of 156
5.15.1
Remote Memory Selection
Memory selection is accomplished by closing pins 1, 9 and 8 as a binary value. The
corresponding memory selected is shown in the table below.
Note:
Only the first seven memory locations can be selected using the remote input
interface. Make sure any test programs are stored in these first seven memories out
of the 50 available memories.
Memory
PIN 1
PIN 9
PIN 8
OFF
OFF
OFF
M1
OFF
OFF
ON
M2
OFF
ON
OFF
M3
OFF
ON
ON
M4
ON
OFF
OFF
M5
ON
OFF
ON
M6
ON
ON
OFF
M7
ON
ON
ON
Table 6-11: Remote Input Memory Selection
5.15.2
Remote Output Interlock
The CFS300 Series is equipped with a Remote Output Interlock feature. Remote Interlock
utilizes a contact closure to enable the power source’s output. If the Remote Interlock
contacts are open, the output of the instrument will be disabled. Remote Interlock is also
referred to as a remote system lockout, utilizing “fail when open” logic.
If the Remote Interlock contacts are open and the TEST button is pushed, a pop-up message
will be displayed on the screen for two seconds.
If the Remote Interlock contacts are opened during a test, the pop-up message will be
displayed and the test will abort. The hardware has been configured to
provide the interlock connections on pins 4 and 5 of the Remote
Interface Input port. The instrument can still be used without the
external interlock device as long as the Interlock Disable Key is plugged
into the Remote Interface port. Refer to Section 6.4, “Ship Kits”.
Note:
If there is nothing connected to the Remote Interface, the instrument will not
perform any tests or provide any output.