CHAPTER 3
MAINFRAME OPERATION
Entire Contents Copyright
2014 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
44M04 LOAD MAINFRAME OPERATIONS MANUAL
Page 35 of 93
44M00
3.6.6
AUTO-SEQUENCE – TEST MODE
To execute an existing auto-sequence, proceed as follows:
1-
Pressing the TEST button will cause the TEST switch to illuminate and the LCD to show
the last selected Auto-sequence number (F1 to F9).The numbered STATE buttons (1-9)
are used to change the Auto-sequence number (F1 ~ F9). Once the desired auto-sequence
has been selected press ENTER to start the test.
2-
The LCD will display "SXX", where "XX" is the actual STEP being presently tested. If
during a given test step the values measured are outside the preset limits (and the NG
Function has been enabled) then the LCD will flash "NG" and the test is suspended.
3-
The user can press the ENTER button to continue the remaining test steps. Alternatively,
the EXIT button may be used to leave the test mode.
4-
If all the test steps are OK, the LCD displays "PASS". If the unit under test fails any of the
test steps the mainframe’s LCD will show "FAIL".
5-
If the buzzer is set to ON, the PASS result is also accompanied by a single buzzer call out.
If the test shows FAIL the buzzer will sound twice.
6-
When the test is completed, the user can press the ENTER key again to start the test again.
Alternatively, the EXIT button may be used to leave the test mode.
Example 1:
Once editing of the 16-step test is completed, press the TEST key. The unit will then
automatically run through the test steps S01 to S16 in order. If all tests steps have been
completed then the LCD will show PASS.