5-12
Veriti™ Thermal Cycler User Guide
Chapter 5
Utilities
1. In the Main Menu screen, touch
Tools Menu
.
2. In the Tools Menu screen, touch
Run
Heated Cover Verification
.
3. Follow the instructions on the screen to perform the test.
4. At the end of the test, touch:
•
to print the results.
•
Save
to save the results to the USB drive.
The results are saved to the file heatedCover.txt.
5. If the test fails, repeat the procedure. Ensure the:
• Probe plate is in the correct location. Probe 1 should be in well A5 during
this test.
• Cables are not damaged (no nicks, cuts or crimps) and are securely seated
in the digital thermometer and in the instrument.
• Heated cover is closed.
• Results are saved or printed.
6. If the test fails again, contact Applied Biosystems Technical Support.
Running the Temperature Verification
Use this test to verify that the average temperature for the sample block(s) is within
±
0.25
°
C of the set point temperature.
Required
Materials
The appropriate Temperature Verification Kit for your Veriti Thermal Cycler model
(see table above).
Procedure
This test takes 25 to 30 minutes to complete, but you can cancel before it is complete.
PHYSICAL INJURY HAZARD.
Hot Surface. Use care when
working around the heated cover and sample block(s) to avoid being burned.
1. In the Main Menu screen, touch
Tools Menu
.
2. In the Tools Menu screen, touch
Run
Temperature Verification
.
3. Follow the instructions on the screen to perform the test.
4. At the end of the test, touch:
•
to print the results.
•
Save
to save the results to the USB drive.
The results are saved to the file bcalevents.txt.
5. If the test fails, repeat the procedure. Ensure the:
• Probe is placed in the correct wells during the test.
• A drop of mineral oil is in each of the test wells.
• Cables are not damaged (no nicks, cuts or crimps) and are securely seated
in the digital thermometer and in the instrument.
• Results are saved or printed.
6. If the test fails again, contact Applied Biosystems Technical Support.
Summary of Contents for Veriti
Page 1: ...Applied Biosystems Veriti Thermal Cycler User Guide ...
Page 2: ......
Page 3: ...Applied Biosystems Veriti Thermal Cycler User Guide ...
Page 8: ...vi Veriti Thermal Cycler User Guide ...
Page 12: ...x Veriti Thermal Cycler User Guide Preface ...
Page 78: ...3 26 Veriti Thermal Cycler User Guide Chapter 3 Performing a Run ...
Page 82: ...4 4 Veriti Thermal Cycler User Guide Chapter 4 Troubleshooting ...
Page 96: ...5 14 Veriti Thermal Cycler User Guide Chapter 5 Utilities ...
Page 106: ...A 4 Veriti Thermal Cycler User Guide Chapter A Instrument Specifications ...
Page 118: ...C 8 Veriti Thermal Cycler User Guide Chapter C Predefined Run Methods ...
Page 123: ......