Veriti™ Thermal Cycler User Guide
Chapter 5
Utilities
Running the Cycle Performance Test
Use this test to verify that the sample block(s) are heating and cooling properly and
that the PCR cycling function operates properly.
Required
Materials
The appropriate equipment for your instrument:
• Veriti 96-Well
Fast
Thermal Cycler: MicroAmp
™
Fast
Optical 96-Well
Reaction Plate with Barcode (PN 4346906)
• Veriti 96-Well Thermal Cycler: MicroAmp
™
Optical 96-Well Reaction Plate
(PN N8010560)
• Veriti 384-Well Thermal Cycler: MicroAmp
™
Optical 384-Well Reaction Plate
(PN 4309849)
• Veriti 60-Well Thermal Cycler: GeneAmp
®
Thermal Isolation Frame
(PN 4308927)
Procedure
This test takes 10 minutes to complete.
1. Place the reaction plate or thermal isolation frame in the instrument.
2. In the Main Menu screen, touch
Tools Menu
.
3. In the Tools Menu screen, touch
Run
Cycle Performance Test
.
4. Follow the instructions on the screen to perform the test. The results are
displayed at the end of the test.
5. At the end of the test, touch:
•
to print the results.
•
Save
to save the results to the USB drive.
The results are saved to the file CyclePEvents.txt.
6. If the test fails, repeat the procedure. Ensure that the heated cover is closed.
7. If the test fails again, contact Applied Biosystems Technical Support.
Summary of Contents for Veriti
Page 1: ...Applied Biosystems Veriti Thermal Cycler User Guide ...
Page 2: ......
Page 3: ...Applied Biosystems Veriti Thermal Cycler User Guide ...
Page 8: ...vi Veriti Thermal Cycler User Guide ...
Page 12: ...x Veriti Thermal Cycler User Guide Preface ...
Page 78: ...3 26 Veriti Thermal Cycler User Guide Chapter 3 Performing a Run ...
Page 82: ...4 4 Veriti Thermal Cycler User Guide Chapter 4 Troubleshooting ...
Page 96: ...5 14 Veriti Thermal Cycler User Guide Chapter 5 Utilities ...
Page 106: ...A 4 Veriti Thermal Cycler User Guide Chapter A Instrument Specifications ...
Page 118: ...C 8 Veriti Thermal Cycler User Guide Chapter C Predefined Run Methods ...
Page 123: ......