Chapter 9 Troubleshooting
9-18
Applied Biosystems
9
Partially resolved peaks
not detected
Mass resolution set too
high to detect average
mass
Decrease Mass
Resolution setting. See
Section 3.2.2, Strategy for
Voyager Peak Detection.
%Base Peak Intensity not
adjusted correctly
Adjust. See Section 3.2.2,
Strategy for Voyager Peak
Detection.
Table 9-11 Peak Detection and Labeling Troubleshooting—Voyager Only
Symptom
Possible Cause
Action
Table 9-12 Charge State and Isotope Determination Troubleshooting—Mariner
Only
Symptom
Possible Cause
Action
Known isotope labeled
with incorrect charge state
(too low)
Max Charge State
parameter set too low.
See example in “Max
Charge State set too low”
on page 3-33.
Set Max Charge State
correctly. See “Peak
Processing parameters
(spectrum data only)” on
page 3-26.
Summary of Contents for Data Explorer 4 Series
Page 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Page 10: ...Table of Contents x Applied Biosystems ...
Page 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Page 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Page 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Page 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Page 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Page 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Page 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Page 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Page 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...