Chapter 8 Viewing Voyager PSD Data
8-12
Applied Biosystems
8
8.3.2 Calibrating
This section includes:
•
Calibrating
•
Matching peaks automatically
•
Selecting peaks manually
•
Solving and plotting
•
Applying new constants to the data file
•
Selecting calibration peaks for optimum mass accuracy
Calibrating
To calibrate a PSD spectrum:
1.
Click the Spectrum window to activate it. Select the
spectrum of interest.
NOTE: You cannot calibrate a result spectrum or a
composite spectrum you have accessed from the
Processing History command on the Display menu. For
more information, see Section 2.4.7, Recalling and
Rearranging Traces (Processing History).
2.
From the Peaks menu, select Peak Label, then select the
Mass Label Type (peak apex or peak centroid) to use for
calibration. Click OK.
3.
From the Process menu, select Mass Calibration, then
select PSD Calibration.
The PSD Processing dialog box is displayed with the
Calibration tab selected (Figure 8-5).
Hint: You can also access this dialog box by selecting
PSD Processing from the Process menu, then clicking
the Calibration tab.
Summary of Contents for Data Explorer 4 Series
Page 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Page 10: ...Table of Contents x Applied Biosystems ...
Page 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Page 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Page 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Page 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Page 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Page 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Page 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Page 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Page 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...