Adjusting the Baseline
Data Explorer
™
Software User’s Guide
5-51
5
Peak Width
(at half height)
(continued)
Set Peak Width according to the data you are correcting:
•
For best results, set to the peak width at half height
of the narrowest peak. However, smaller peak width
values increase processing time.
•
If peak width varies across the spectrum, set to the
average peak width.
•
If peaks are narrow relative to the baseline region
you are correcting, set a larger value (for example,
10 times the width of widest peak) to increase
processing speed.
•
If the baseline changes sharply across the spectrum,
set a smaller value closer to the narrowest peak
width.
•
If the baseline is broad or gently sloping, set a larger
value, for example 10 to 20 times the width of the
actual peak width.
To determine the number of points across a peak, change
the trace display from Line to Vertical Bars (select
Graphic Options from the Display menu, click Graph #
Setup and change Line Type). Each vertical bar
represents one data point. For more information, see
Section 1.4, Customizing the Data Explorer Window.
(continued)
Parameter
Description/Specifies
Summary of Contents for Data Explorer 4 Series
Page 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Page 10: ...Table of Contents x Applied Biosystems ...
Page 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Page 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Page 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Page 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Page 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Page 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Page 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Page 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Page 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...