Chapter 5 Examining Spectrum Data
5-22
Applied Biosystems
5
5.3.4 Reverting to Instrument Calibration
The Revert to Instrument Calibration function does the
following:
•
Mariner data—Reapplies the original calibration
constants used to acquire the data.
•
Voyager data—Applies default calibration to the data,
regardless of whether default or external calibration is
used to acquire the data.
To revert the calibration:
1.
From the Process menu, select Mass Calibration, then
select Revert to Instrument Calibration.
The spectrum is recalibrated with the original
calibration constants used during acquisition (Mariner
data) or default calibration (Voyager data) and
displayed with an MC trace label. The original and new
constants are displayed in the Output window.
2.
To save the reverted calibration with the data file, select
Mass Calibration from the Process menu, then:
If you are
calibrating
Select
The following occurs
Mariner
data
(including
MS Method
data)
Apply Calibration
All spectra in the data file are calibrated and
displayed with an MC trace label. The
calibration constants are saved with the data
file. Each spectrum in the data file is
calibrated when displayed.
Summary of Contents for Data Explorer 4 Series
Page 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Page 10: ...Table of Contents x Applied Biosystems ...
Page 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Page 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Page 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Page 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Page 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Page 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Page 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Page 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Page 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...