![Applied Biosystems Data Explorer 4 Series User Manual Download Page 167](http://html.mh-extra.com/html/applied-biosystems/data-explorer-4-series/data-explorer-4-series_user-manual_2955117167.webp)
Process that Occurs During Peak Detection, Centroiding, and Integration
Data Explorer
™
Software User’s Guide
3-69
3
After peak
detection
After peaks are detected:
•
Centroid mass is calculated for spectral data, then
modified by Gaussian peak fitting, if it is selected.
•
Chromatographic and spectral peaks are integrated.
•
Peak lists are generated for chromatograms and spectra.
Centroiding
The Data Explorer software calculates the peak centroid for
spectral data by:
•
Drawing a projected centroid baseline at the
percentage of the peak height specified by the
Centroiding% entered. Centroiding is measured
between the top of the peak and 0. For example, a
centroiding% of 10 uses the top ten percent of the peak,
where peak height is determined from 0.
•
Searching to the left and right of the apex for data
points that bound the projected centroid baseline, then
interpolating the centroid baseline.
•
Calculating the intensity-weighted area of the peak
from the left to the right intersection of the centroid
baseline, then dividing by the area of the same
region.
Summary of Contents for Data Explorer 4 Series
Page 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Page 10: ...Table of Contents x Applied Biosystems ...
Page 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Page 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Page 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Page 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Page 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Page 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Page 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Page 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Page 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...