Chapter 7
Verifying the Instrument Performance
Analyzing the TaqMan
®
RNase P Plate Data
94
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Analyzing the TaqMan
®
RNase P Plate Data
1.
In the SDS software, select
Analysis
Analysis Settings
.
2.
In the Analysis Settings dialog box:
a.
Select
Auto Ct.
b.
Click .
3.
In the SDS software, click
(or select
Analysis Analyze
).
The SDS software analyzes the run data and
displays the results in the Results tab.
4.
Access the results. In the plate document:
a.
Select the
Results
tab.
b.
Select the
Amplification Plot
tab.
Select
b
a
Select
a
b
Summary of Contents for 7300