Clean the polymer delivery pump (PDP) and lower polymer block
• Do not expose the polymer blocks to organic solvents.
• Do not use sharp or pointed instruments to remove dried polymer from the
polymer blocks.
• Do not use water >40°C to clean the polymer blocks.
• Clean the exterior 1 time per week, when you replenish polymer.
• Flush the polymer delivery pump water trap 1 time per week.
• Clean the polymer delivery pump chamber, channels, and tubing 1 time per
month.
Parts of the
polymer delivery
pump
Guidelines
Frequency
Chapter 9
Maintenance
Clean the polymer delivery pump (PDP) and lower polymer block
9
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
121