Dye Set J6 created from Matrix Standard Set DS-36
To export spectral calibration results:
1.
Click Export Spectral Calibration Results.
2.
Specify an export file name and location, then click Save.
The export file contains the following results:
• Capillary Number
• Condition Number
• Scan Number
• Borrowed From Capillary
• Quality Value
• Peak Height
• Reason For Failure
• Run From Injection
Note:
Spatial and spectral calibration reports include the date on which a capillary
array is installed for the first time on the instrument. Install standard reports use the
most recent install date if a capillary array was removed and re-installed on the
instrument.
1.
Click View Report.
2.
(Optional) In the Report screen, click toolbar options to manipulate the report.
Place the mouse pointer over an item for a description of the item.
3.
To print the report, click Print.
Export spectral
calibration results
View and print a
calibration report
Chapter 6
Run calibrations and install checks
Run a spectral calibration
6
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
119