Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 8
Temperature non-uniformity test
12. When you complete all measurements:
a. Press
F5
(Cancel).
b. Remove the 0.1-mL probe assembly from the sample blocks.
c. Power off the digital thermometer, then clean off the oil.
IMPORTANT!
Make sure the sample blocks are at room temperature (~25 °C)
before removing the reaction plates.
Evaluating the
results
When the GeneAmp
®
9800 completes the temperature non-uniformity test, the TNU
Performance screen opens. Use the table below to evaluate the results.
If the...
Then...
Temperature of the
sample block wells is
uniform
“Pass” appears after each setpoint temperature.
Temperature variation
of the sample block
wells exceeds
performance
specifications
“Fail” appears after the setpoint temperature(s) for which the
test failed.
• If the test fails, repeat the procedure to make sure that the
thermometer was not misread or that data-entry errors were
not made.
• If the test fails again, contact Applied Biosystems Technical
Support. See
“How to obtain support” on page viii
.
TNU Performance
TNU at 95°C is xx.xx - Pass
TNU at 55°C is xx.xx - Pass
F1
F3
F4
F5
F2
Cancel
TNU Performance
TNU at 95°C is xx.xx - Fail
TNU at 55°C is xx.xx - Fail
F1
F3
F4
F5
F2
Cancel