CSP-CSPS-8000/B Section 1 Page 25 of 30
CONNECTORS
MAINS INPUT
200
– 250 VAC mains input. Connection details are supplied elsewhere in this manual.
REMOTE
Allows operator to remotely control CSP-S.
HV OUTPUT
4 pin proprietary connector. Using our HV2 plug, the large connector pins are proof
tested to 6000 volts and can operate with current pulses up to 6000 amps. The two small
pins are used as interlock pins. If these are left open circuit, the CSP-D will not be able to
enable the high voltage (INTERLOCK fault).
TRIGGER INPUT
The trigger input is connected via a BNC socket.
REAR PANEL
FUSES
The rear panel contains 2 fuse holders, and these are the only things we recommend the
operator checks if he has not received factory training. The fuses are rated as follows:-
CONTROL UNIT
F1
3A Antisurge 1¼"
(240V)
7A Antisurge 1¼"
(110V)
F2
3A Antisurge 1¼"
(240V)
7A Antisurge 1¼"
(110V)
ENERGY UNIT
F3
3A Antisurge 1¼"
(240V)
7A Antisurge 1¼"
(110V)
FAN FILTERS
The rear panel of each of the units has 2 fold-down handles to ease installation, and they
also have 2 fan filters for the cooling air intake. These filters should be periodically
cleaned to ensure maximum air flow. This can be done with a small stiff brush.