CSP-CSPS-8000/B Section 1 Page 23 of 30
O
PERATOR
C
ONTROLS AND
I
NDICATORS
MAINS POWER switch
Double pole switch that also acts as an overcurrent circuit breaker. It is situated by the
mains input connector. The switch has standard O I positions.
EMERGENCY STOP switch
In the event of an emergency the unit can be shut down. This switch shuts off the control
logic of the charger thus disabling the system. Rotate to reset.
HV OFF/RESET switch
Needs to be pressed if fault occurs, or if someone has tampered with the controls. This
button also operates as the HV OFF button. Press to switch OFF.
HV ON
switch
Switches in capacitors and HV PSU, (note delays on switching), switch illuminates when
the high voltage is switched on operated in conjunction with the HV ENABLE key switch,
thus to switch HV ON, HV ENABLE needs to be enabled simultaneously.
MANUAL KEY switch
Triggers CSP unit when pressed. Also illuminates when an external key pulse is detected
KEY POLARITY switch
Used to select either:
+ve key pulse (3.5
– 20 volts, opto isolated triggers 0.5 mS from rising edge), or
contact closure (triggers 0.5 mS from falling edge or closure - not isolated).
LOCAL / REMOTE switch
Controls whether an external remote box can be used, as marked on the front panel.
Note that the unit can be switched off from the front panel and the remote at any time.
300J LIMIT switch (Located on front panel)
Used to limit the output power to 300J when latched in and illuminated. When not
engaged the output power is user selectable.