AT381x User's Guide
84
10.14.2
CORRection:SHORt
The CORRection:SHORt command execute all presetted SHORT correction data measurement
points.
Command Syntax
CORRection:SHORt
Example
SEND> CORRection:SHOR
RET>
short
10.14.2.1
CORRection:SHORt:STATe
The CORRection:SHORt:STATe command sets the SHORT correction function to ON or OFF.
Command Syntax
CORRection:SHORt:STATe {ON,OFF,1,0}
Parameter
Where, {ON,OFF,1,0} is:
ON, 1
When the function is ON
OFF,0
When the function is OFF
Example
SEND> CORR:SHOR:STATe ON
Query Syntax
CORRection:SHOR:STATe?
Query Response
{on,off}
Example
SEND> CORR:SHOR:STATe?
RET>
on
10.14.3
CORRection:SPOT:FREQuency
The CORRection:SPOT:FREQuency command sets the frequency point for the specified
frequency point correction.
Command Syntax
CORRection:SPOT:FREQuendy <value>
Parameter
Where, <value> is:
value
NR1,NR2 or NR3:Frequecny value.
A suffix multiplier can be used with this command. But the unit “Hz” cannot be added.
Example
SEND> CORR:SPOT:FREQ 1k
SEND> CORR:SPOT:FREQ 10k
Query Syntax
CORRection:SPOT:FREQuency?
Query Response
<NR3>
Example
SEND> CORR:SPOT:FREQ?
RET>
1.03
10.14.4
CORRection:SPOT:OPEN
This command executes the OPEN correction data measure for the specified frequency
correction.
Command Syntax
CORRection:SPOT:OPEN
Example
SEND> CORR:SPOT:OPEN
10.14.5
CORRection:SPOT:SHORt
This command executes the SHORT correction data measure for the specified frequency
correction.
Command Syntax
CORRection:SPOT:SHORt
Example
SEND> CORR:SPOT:SHOR