AT381x User's Guide
82
Command Syntax
LIST:BAND <n>,<point value>,{A,B,OFF},<low>,<high>
Parameter
Where, <n>,<point value>,{A,B,OFF},<low>,<high> is:
n
NR1(1 to 10): List sweep point
<point value>
sweep point value (frequency value or signal level voltage value)
A,
Uses the primary parameter as the limit parameter.
B,
Uses the secondary parameter as the limit parameter.
OFF
Turn off the List Sweep's comparator function
<low>
NR1,NR2 or NR3: low limit value
<high>
NR1,NR2 or NR3: high limit value
Note: The suffix multipliers can be used with this command. But the suffix units CANNOT
be added.
Example
SEND> LIST:BAND 1,1k,A,1n,2n
SEND> LIST:BAND 2,10k,A,1E-9,2E-9
Query Syntax
LIST:BAND? <n>
Parameter
Where,<n> is:
n
NR1(1 to 10): List sweep point
Query Response
{on,off},<point value>,{A,B,-},<NR3:low>,<NR4:high>
Example
SEND> LIST:BAND? 1
RET>
on,1.03,A,1.000000E-9,2.000000E-9
10.14
CORRection Subsystem
The CORRection subsystem command group sets the correction function, including the
OPEN, SHORT and LOAD correction settings.
NOTE:
The CORRection subsystem CANNOT work in <LIST MEAS> page.
Figure 10-9
CORRection Subsystem Command Tree