55
System Configurations
Soft key
Function
OFF
Will not return error code
ON
Return an error code
NOTE
This setting and function is valid only under the SCPI protocol.
6.1.11
SCPI [RESULT] Setting
When the result sending function is set to automatic, the data for each measurement is
automatically sent to the host.
To set up the result send:
Step 1
Press the [Meas] or [Setup] key and then press [SYSTEM] TO enter <SYSTEM CONFIG> page
Step 2
Use the cursor key to select [Result Send] field
Step 3
Use the soft keys to select.
Soft key
Function
FETCH
The test result will be sent back by sending "fetch?" instruction.
ON
The test result will be sent back after measuring.
NOTE
This setting and function is valid only under the SCPI protocol.
6.1.12
[DATA BUFFER]
Set the maximum data buffer value for the data logging function. The AT381x can set up to
10000 sets of buffer data. After the cache setting value is reached, the data record will stop.
This data can be saved into the external USB disk.
Please refer to the [LOG] field in the <MEAS DISPLAY> page.
6.1.13
Restore to [DEFAULT SET]
[DEFAULT SET] setting option allow user settings to be restored to factory settings.
6.2
System Information Page
Press [Meas] or [Setup] key, press the [SYSTEM] key at bottom to enter the <SYSTEM
CONFIG> page, press the soft key to select [SYSTEM INFO].
There are no configurable options in the system information page
Figure 6-2
< SYSTEM INFORMATION> page