AT381x User's Guide
30
Soft key
Function
START LOG
Start a new measurement data record
Step 4
Press [START LOG] soft key to start recording data to internal buffer of the instrument.
Options after the start of recording
Soft key
Function
STOP
The current record is stopped.
SAVE & STOP
The current record is stopped and the data is saved to a USB disk file
CLEAR & STOP
The current record is stopped, the buffer is cleared, but the data is not
saved to the USB disk.
Step 5
If press the soft key [SAVE], [SAVE & STOP] or [CLEAR & STOP], the current recording is
terminated.
Or wait for the buffer to be full, press [SAVE to USB] button to save data to the USB disk.
Soft key
Function
SAVE to USB
Save data to a USB disk file
CLEAR BUFFER
The buffer is cleared and the current data is invalid.
Note:
Before using the saved data, insert the USB disk on the interface of front panel.
The saved file is in the DATA subfolder in the same folder as the USB disk. The file
name is: 001.CSV For example: F:\AT3818\DATA\001.CSV
Up to 1000 files can be saved (from 001.CSV to 999.CSV)
The buffer size can be modified in the [DATA BUFFER] field on the [SYSTEM
CONFIG] page. (maximum 10000 data)
4.2
<BIN MEAS> Page
Press [Meas] key and use soft key to enter the [BIN MEAS] page.
Figure 4-2
<BIN MEAS> page
Setting Bar: The setting area of bin number display page is roughly same as the [Meas Display]
page. These settings include the following 8 items:
Function [FUNC]
Range [RANGE]
Frequency [FREQ]
Trigger [TRIG]
Level [LEVEL]
Speed [SPEED]
Comparator [COMP]
Auxiliary bin [AUX]