AT381x User's Guide
24
4.
[Meas] Page
This section includes the following information:
•
MEAS DISPLAY page
•
BIN MEAS page
•
BIN COUNT page
•
LIST SWEEP page
4.1
<MEAS DISPLAY> Page
Press [Meas] key to enter <MEAS DISPLAY> page.
<Meas Display> page mainly highlights measurement results and displays the current sorting
result in small characters.
The following 6 common functions can be set on this page
• LOG – Record test data in a USB disk file
• FUNC - Measurement function
• RANGE - Measurement range, range number and automatic/manual test.
• FREQ - Measurement frequency
• TRIG - Trigger Setup
• LEVEL - Test signal level
• SPEED – Test speed
The primary and secondary test results are displayed in this area in large characters, monitor
area is displayed in small characters.
In the lower part of the screen, some measurement-related settings are displayed in this area,
and the currently tested comparator results are also displayed here.
Figure 4-1
<MEAS DISPLAY> Page
4.1.1
Measurement Function [FUNC]
AT381x simultaneously measures four components of the complex impedance (parameters) in
a measurement cycle. These include primary parameter, secondary parameter and two
monitor parameters.