121
Modbus
(
RTU
)
Command
Returns FFFF, indicating that the correction failed
Note:
It takes time to perform correction of the point frequency. Please delay the time in the
software to return the data.
12.6.5
Spot Frequency Short-Circuit Correction [5028H]
By writing the point frequency 0001~0003 to the register 5028, the AT381x will begin to
perform an open-circuit point frequency correction.
After clearing is complete, you can get the clear status by reading the 5028 register:
0000
Correction success
FFFF
Correction fail
Write
Please write the point frequency to the 5028 register: 0001~0003, and execute the
corresponding point frequency short-circuit correction.
Send: 01 10 5028 0001 02 0001 31BD
Response: 01 10 5028 0001 90C1
Read
During the execution of correction, can determine whether correction is completed by
reading the register data.
Send
:
01 03 5028 0001 1502
Response
:
01 03 02 FFFF B9F4
Returns FFFF, indicating that the correction failed
Note:
It takes time to perform correction of the point frequency. Please delay the time in the
software to return the data.
12.7
System Setup
12.7.1
Instrument version number [0000H]
Read-only register, register [0000] ~ [0003] return the version number of the instrument:
Read
1
2
3
4
5
6
7
8
01
03
00
10
00
02
4A
6D
read
register
number of
registers
CRC
Response
:
1
2
3
4
5
6
7
8
9
01
03
04
43
37
30
30
4A
6D
byte
data ASCII: C700
CRC
The version number is ASCII value: for example, 43 37 30 30 = C700