AT381x User's Guide
120
12.6.3
Spot frequency Correction Setting [5010H-5015H]
The AT381x point frequency can be set by the register [point frequency 1:5010~5011], [point
frequency 2:5012~5013] and [point frequency 3:5014~5015], where set to 00000000 means
the point frequency is off.
Write
Point frequency 1
:
1kHz (Single precision floating point number
:
44 7A 00 00)
1
2
3
4
5
6
7
8
9
10
11
12
13
01
10
50
10
00
02
04
44
7A
00
00
3B
89
write
register
number of
registers
byte
data
CRC
Response
:
1
2
3
4
5
6
7
8
01
10
50
10
00
02
51
0D
register
number of
registers
CRC
Read
1
2
3
4
5
6
7
8
01
03
50
10
00
02
CF
1A
read
register
number of
registers
CRC
Response
:
1
2
3
4
5
6
7
8
9
01
03
04
44
7A
00
00
CF
1A
byte
data 1000
CRC
12.6.4
Spot Frequency Open Circuit Correction [5020H]
By writing the point frequency 0001~0003 to the register 5020, the AT381x will begin to
perform an open-circuit point frequency correction.
After clearing is complete, you can get the clear status by reading the 5008 register:
0000
Correction success
FFFF
Correction fail
Write
Please write the point frequency to the 5020 register: 0001~0003, and execute the
corresponding point frequency open circuit correction.
Send: 01 10 5020 0001 02 0001 30F5
Response: 01 10 5020 0001 1103
Read
During the execution of correction, can determine whether correction is completed by
reading the register data.
Send
:
01 03 5020 0001 94C0
Response
:
01 03 02 0000 B844