![Applent Instruments Anbai AT8612 User Manual Download Page 55](http://html.mh-extra.com/html/applent-instruments/anbai-at8612/anbai-at8612_user-manual_2954893055.webp)
55
55
SCPI Command
Inc.
7.7
FETCH subsystem
Table 7-3 FETCh Subsystem Command Tree
FETCH
:MEASURE
:CURRENT
:VOLTAGE
:POWER
:RESISTANCE
FETCh subsystem is used to retrieve measurement data
7.7.1
FETCH:MEASURE
FETCH:MEASURE query all measurement data
Query Syntax
:
FETCH:MEASURE
Query Response
: {float, float, float, float}
Group 1's data: current value
Group 2's data: voltage value
Group 3's data: power value
Group 4's data: resistance value
Example
:
SEND
>
fetch:meas
RECEIVE> 1.0000,120.00,300.00,4000.0
7.7.2
FETCH: CURRENT
FETCH: CURRENT command is used to retrieve the actual current value currently, unit
A.
Query Syntax
:
FETCH:CURRENT
Query Response
:
<
float
>
Example
:
SEND
>
fetch:curr
RET
>
1.0000
7.7.3
FETCH: VOLTAGE
FETCH:VOLTAGE
command
is used to retrieve the actual voltage value currently, unit
V.
Query Syntax
:
FETCH:VOLTAGE
Query Response
:
<
float
>
Example
:
SEND
>
fetch:volt
RET
>
120.00
7.7.4
FETCH:POWER
FETCH:POWER
command
is used to retrieve the actual power value currently, unit W
Query Syntax
:
FETCH:POWER
Query Response
:
<
float
>
Example
:
SEND
>
fetch:POW
RET
>
10.000
Summary of Contents for Anbai AT8612