19
19
Start up
2.5
Protection Features
To maximize protect the measured power supply, AT861x includes five protection features.
2.6
Over Voltage
Load input will be turned OFF immediately if input voltage exceeds the 105%*V-MAX,
load overvoltage alerts, voltage value glitters.
When input voltage exceed 110%, load input is turned off and display [
OV
], the buzzer
beeps.
WARNING
1.
Don't input voltage that exceeds rated voltage, over voltage protection is dangerous for load.
2.
NOTE: In the setup page, over voltage protection only provide [
OV
] and buzzer beeping.
3.
Suggestion: Make sure there’s no over voltage happened before enter all setup pages.
4.
Over voltage, the load input will be turned off under any mode.
2.7
Over Current
In the CV mode, when input current exceeds I-MAX, load over-current alerts, current
value glitters and the buzzer beeps. Once input current exceeds 102%, load input is turned
off and display [
OC
].
When work in the CR or CC and CP mode, input current is ascending continuously and
Input
Voltage(V)
Input Current (I)
Rated Power
Rated
Voltage
Rated Current
Input
Voltage(V)
Input Current (I)
Rated
Voltage
Rated Current
Current Limit
Power Limit