background image

9. Maintenance, calibration, and
troubleshooting

NOTE:
All factory repairs and recalibrations require a returned material authorization (RMA) and
completion of the “Statement of Product Cleanliness and Decontamination” form. Refer to
the

Assistance

page at the end of this manual for more information.

9.1 Maintenance

Moisture or debris on the filters (diffuser for upward-looking pyranometer, glass window for
downward-looking pyranometer, silicon windows for pyrgeometers) is a common cause of
errors. Remove dust or organic deposits by using water or window cleaner and a soft cloth or
cotton swab. Remove salt deposits by using vinegar and a soft cloth or cotton swab.

CAUTION:
Never use an abrasive material or cleaner on the diffuser.

9.2 Calibration

The Clear Sky Calculator (

www.clearskycalculator.com

) can be used to determine the need for

pyranometer recalibration. It determines total shortwave radiation incident on a horizontal
surface at any time of day at any location in the world. It is most accurate when used near solar
noon in spring and summer months, where accuracy over multiple clear and unpolluted days is
estimated to be ±4% in all climates and locations around the world. For best accuracy, the sky
must be completely clear, as reflected radiation from clouds causes incoming radiation to
increase above the value predicted by the clear sky calculator. Measured values of total
shortwave radiation can exceed values predicted by the Clear Sky Calculator due to reflection
from thin, high clouds and edges of clouds, which enhances incoming shortwave radiation. The
influence of high clouds typically shows up as spikes above clear sky values, not a constant offset
greater than clear sky values.

To determine recalibration need, input site conditions into the calculator and compare total
shortwave radiation measurements to calculated values for a clear sky. If sensor shortwave
radiation measurements over multiple days near solar noon are consistently different than
calculated values (by more than 6 %), the sensor should be cleaned and re-leveled. If

SN500SS Net Radiometer

13

Summary of Contents for SN-500-SS

Page 1: ...Revision 08 2021 Copyright 2018 2021 Campbell Scientific Inc...

Page 2: ...ong cables 12 9 Maintenance calibration and troubleshooting 13 9 1 Maintenance 13 9 2 Calibration 13 9 3 Troubleshooting 14 Appendix A Importing Short Cut code into CRBasic Editor 15 Appendix B SDI 12...

Page 3: ...c redundancy check aMC and aCC 21 B 1 9 Stopping a measurement command 21 B 1 10 Send data command aD0 aD9 21 B 1 11 Continuous measurement command aR0 aR9 22 B 1 12 Extended commands 22 B 2 SDI 12 tr...

Page 4: ...TAND the Safety section at the back of this manual l Care should be taken when opening the shipping package to not damage or cut the cable jacket If damage to the cable is suspected consult with Campb...

Page 5: ...procedure also shows using Short Cut to program the SN500SS 1 Open Short Cut and click Create New Program 2 Double click the data logger model 3 In the Available Sensors and Devices box type SN500SS...

Page 6: ...wiring the sensor 5 Repeat steps three and four for other sensors 6 In Output Setup type the scan rate meaningful table names and Data Output Storage Interval NOTE Because of the delays using SDI 12 C...

Page 7: ...he sensor is connected to the data logger check the output of the sensor in the data display in LoggerNet RTDAQ or PC400 to make sure it is making reasonable measurements 5 Overview The SN500SS is a f...

Page 8: ...on accounts for a larger proportion of net radiation during the day when the sun is shining Longwave radiation contributes to net radiation during the day and at night Net shortwave radiation is the d...

Page 9: ...1 s SDI 12 data transfer rate detector response times are 0 5 s Heater current drain at 12 VDC 63 mA heaters on communication enabled 1 5 mA heaters off communication enabled 0 6 mA heaters off commun...

Page 10: ...complex the files that Short Cut creates are a great source for programming code to start a new program or add to an existing custom program NOTE Short Cut cannot edit programs after they are importe...

Page 11: ...ample if the instrument is mounted 6 m above the surface 99 of the input of the lower sensors comes from a circular area with a 60 m radius To avoid shading or reflection effects and to promote spatia...

Page 12: ...nd cable connected 2 Attach the mounting bracket to the vertical mounting pipe or CM200 series crossarm using the provided U bolt FIGURE 7 3 p 10 If mounted to a vertical pipe ensure that the pipe doe...

Page 13: ...veling of the sensor using the sensor bubble level 5 Tighten the four screws on top of the mounting bracket to properly secure the support arm so that it does not rotate FIGURE 7 3 p 10 CAUTION The fo...

Page 14: ...the values are ready Rather the data logger picks up the data with the D command on the next pass through the program Another measurement request is then sent so that data is ready on the next scan N...

Page 15: ...or incoming LW 3 Voltage signal for outgoing LW 4 Sensor body temperature for outgoing LW mV C mV C aM4 aC4 aMC4 or aCC4 Albedo W m 2 aXHON Turn heater on aXHOFF Turn heater off Returns the SDI 12 Add...

Page 16: ...time of day at any location in the world It is most accurate when used near solar noon in spring and summer months where accuracy over multiple clear and unpolluted days is estimated to be 4 in all c...

Page 17: ...fied by the SDI12Recorder instruction 2 Check the voltage to the sensor with a digital voltage meter If a switched 12V terminal is used temporarily connect the red wire to a 12V terminal non switched...

Page 18: ...ted for additional refinement NOTE Once the file is edited with CRBasic Editor Short Cut can no longer be used to edit the program it created 2 To add the Short Cut wiring information into the new CRB...

Page 19: ...Sniffer Mode videos B 1 SDI 12 command basics SDI 12 commands have three components l Sensor address a a single character and the first character of the command Use the default address of zero 0 unles...

Page 20: ...RC CR LF Extended Commands aXNNN a values CR LF 1 Information on each of these commands is given in the following sections B 1 1 Acknowledge active command a The Acknowledge Active command a is used t...

Page 21: ...until verification information is available n the number of values to be returned when one or more subsequent D commands are issued B 1 4 Address query command Command requests the address of the con...

Page 22: ...0 immediately indicates that it will return two values within the next 35 seconds 0 CR LF Within 35 seconds sensor 0 indicates that it has completed the measurement by sending a service request to th...

Page 23: ...sor Y immediately indicates that it will have 6 06 values ready for collection within the next 40 040 seconds ZC The data logger makes a request to sensor Z to start a concurrent measurement Z02010 CR...

Page 24: ...t www sdi 12 org to learn more about how the CRC code is developed B 1 9 Stopping a measurement command A measurement command M is stopped if it detects a break signal before the measurement is comple...

Page 25: ...in the values part of the response to the R command B 1 12 Extended commands Many sensors support extended SDI 12 commands An extended command is specific to a make of sensor and tells the sensor to...

Page 26: ...use their transparent mode The following examples show how to enter transparent mode and change the SDI 12 address of an SDI 12 sensor The steps shown in Changing an SDI 12 address p 23 are used with...

Page 27: ...sponds with its SDI 12 address If no characters are typed within 60 seconds the mode is exited In that case simply type SDI12 again press Enter and type the correct control port number when prompted 1...

Page 28: ...lar to that shown for the CR1000X B 3 References SDI 12 Support Group SDI 12 A Serial Digital Interface Standard for Microprocessor Based Sensors Version 1 4 River Heights UT SDI 12 Support Group 2017...

Page 29: ...ther manufacturer s products that are resold by Campbell Scientific are warranted only to the limits extended by the original manufacturer Refer to www campbellsci com terms warranty for more informat...

Page 30: ...ion or Repair Reference number contact your CAMPBELL SCIENTIFIC regional office Please write the issued number clearly on the outside of the shipping container and ship as directed For all returns the...

Page 31: ...rohibit climbing by other persons Take reasonable precautions to secure tripod and tower sites from trespassers l Use only manufacturer recommended parts materials and tools Utility and Electrical l Y...

Page 32: ...nfo campbellsci fr www campbellsci fr Germany Location Phone Email Website Bremen Germany 49 0 421 460974 0 info campbellsci de www campbellsci de India Location Phone Email Website New Delhi DL India...

Reviews: