DO8032 Luminescent Sensor Cap Installation Instruction
Traditionally, lots of data need to be manually input to a meter in order to finish the replacement of a
luminescent sensor cap for optical DO probes. In this replacement system, we compile all the data and program
into a flash drive. After installing the program, connect the data box with your PC and sensor, and simply press
the Start button to complete replacement.
1. Please check serial number on the sensor cap and flash drive. The numbers should be consistent.
2. Screw off the sensor cap (1), take off the O-ring (2), as shown in the following figure. Then install the new O-
ring and sensor cap, screw it tightly.
3. Insert the flash drive into a Windows-based computer, install the driver program.
Note: for 32bit OS, please run "McphCdcDriverInstallationTool" file in /x86;
for 64bit OS, please run "McphCdcDriverInstallationTool" file in /x64.
4. Run XXX.exe file in the flash drive, connect computer and sensor according to guide (as shown in the following
graph), then press "Start" button. When "Load is OK" displays at the bottom left corner, the replacement is
completed.
Note: If replacement is not completed, other reminders will display at the bottom left. Please check it according
to the reminder or wait a few more minutes if computer has not completed automatic installing driver program.
5. The DO probe should be re-calibrated after replacing sensor cap. When taking out a new sensor cap, if the
sponge in the black sheath is already dried out, please soak the probe (after installing the sensor cap) in tap
water for 48 hours before using.