Apera 400S Series Instruction Manual Download Page 12

 

- 12 - 

4.4.3 pH isothermal measurement principle 

According to the pH isothermal measurement principle, the closer the test sample’s temperature is to the calibration 
solution’s, the higher the accuracy of the measurement. So this principle is recommended to follow when conducting 
tests. 

4.4.4 Restore to factory default 

The instrument has a function to return to factory default setting, which can be set up in P1.5 (refer to section 7.3). 

Returning to factory default setting is to restore the meter to theoretical value (zero potential pH is 7.00, slope is 

100%), and set all the parameters to default settings (see appendix 1). When the meter’s calibration or measurement 

is performing abnormally, users can use this function to let the meter return to factory default mode, and then conduct 

calibration and test again. Please note that this function is irreversible once used. 

 

 

Maintenance of the pH Electrode 

4.5.1. Daily maintenance 

The soaking solution contained in the supplied protective bottle is used to maintain activation in the glass bulb and 

junction. Loosen the capsule, remove the electrode and rinse in distilled or pure water before taking a measurement. 

Insert the electrode and tighten the capsule after measurements to prevent the solution from leaking. If the soaking 

solution is turbid or moldy, replace the KCL solution. (Preparation: 26g KCL dissolved in 100ml pure water). 

The electrode should not be soaked in pure or distilled water, protein solution or acid fluoride solution for long periods 

of time. In addition, do not soak the electrode in organic lipids. 

 

4.5.2. Calibration buffer solution 

For  calibration  accuracy,  the  pH  of  the  standard  buffer  solution  must  be  reliable.  The  buffer  solution  should  be 

refreshed often, especially after heavy use. We recommend 10-15 times of use before replacing the pH buffers. 

 

4.5.3 Protect glass bulb  

The sensitive glass bulb at the front of the combination electrode should not come in contact with hard surfaces. 

Scratches  or  cracks  on  the  electrode  will  cause  inaccurate  readings.  Before  and  after  each  measurement,  the 

electrode should be washed with pure water and dried. If a sample sticks to the electrode or it’s contaminated, the 

electrode should be thoroughly cleaned using a soft brush and then rinsed with pure water. After that, soak it in the 

KCL solution again for 6 hours. 

 

4.5.4 Renew glass bulb 

Electrodes that have been used over a long period of time will become aged. Submerge the electrode in 0.1mol/L 

hydrochloric acid for 24 hours, then wash the electrode in pure water, then submerge it in KCL soaking solution for 

24 hours.  

The method to prepare 0.1mol/L hydrochloric acid: dilute 9mL hydrochloric acid in pure water to 1000mL. 

For serious passivation, submerge the bulb in 4% HF (hydrofluoric acid) for 3-5 seconds, and wash it in 

pure water, then submerge it in the soaking solution for 24 hours to renew it. 

 

4.5.5. Clean contaminated glass bulb and junction (please refer to Table-8) 

 

 

Summary of Contents for 400S Series

Page 1: ...EC400S Portable Conductivity Meter Instruction Manual PH400S Portable pH Meter EC400S Portable Conductivity Meter PC400S Portable pH Conductivity Meter APERA INSTRUMENTS Europe GmbH www aperainst de v...

Page 2: ...ment 13 6 Conductivity Measurement 14 Information regarding the Conductivity Electrode 14 Information regarding Conductivity Calibration 15 Conductivity Calibration 16 TDS Salinity Resistivity Conduct...

Page 3: ...reminder etc GLP data management real time clock display Manual or Auto timing data storage USB data communication The meter adopts advanced digital processing technology intelligently improves the re...

Page 4: ...ty 0 to 200 mS cm including 5 ranges 0 00 to 19 99 S cm 20 0 to 199 9 S cm 200 to 1999 S cm 2 00 to 19 99 mS cm 20 0 to 199 9 mS cm TDS 0 to 100 g l including 5 ranges 0 00 to 9 99 mg l 10 0 to 99 9 m...

Page 5: ...nd Dustproof Dimensions and Weight PH400S EC400S Meter 86x196x33 mm 335 g With case 360x275x80 mm 1 6 kg PC400S Meter 480x360x95 mm 335 g With case 480x360x95 mm 4 2 kg 3 INSTRUMENT DESCRIPTION LCD Di...

Page 6: ...s 3 2 2 Power On Short press to power on LCD displays the measuring mode used last time Backlight turned on for 1 minute 3 2 3 Power Off The meter can only be turned off in measuring mode by short pr...

Page 7: ...d conductivity electrode can be connected When connected the meter will automatically switch to the correspondent measurement mode Display Mode 3 4 1 Stable Reading Display Mode When the measuring val...

Page 8: ...ng the pH Electrode The instrument is equipped with a 201T S 3 in 1 Combination pH Electrode Its built in temperature sensor allows for auto temperature compensation The electrode s housing adopts PC...

Page 9: ...testing strong acid or strong base solutions 4 2 3 How often to calibrate The frequency that you need to calibrate your meter depends on the tested samples condition of electrodes and the requirement...

Page 10: ...lution Stir the solution gently and let stand still in the buffer solution until a stable reading is reached The meter s display will show the recognition process of calibration buffer solution at the...

Page 11: ...suring value being unstable for over 3 minutes 1 Check whether there are bubbles in glass bulb 2 Replace with a new pH electrode pH electrode zero electric potential out of range 60mV or 60mV 1 Check...

Page 12: ...e water The electrode should not be soaked in pure or distilled water protein solution or acid fluoride solution for long periods of time In addition do not soak the electrode in organic lipids 4 5 2...

Page 13: ...de for your meter in order to accurately test purified waters 2 The electrode housing is polycarbonate When using cleaning solutions take cautions on carbon tetrachloride trichlorethylene tetrahydrofu...

Page 14: ...which is corrosion resistant and impact resistant When submerge the conductivity electrode in solution stir the solution briefly to eliminate the air bubbles and improve response and stability 6 1 2...

Page 15: ...to the sample solution to perform 1 point or 2 point calibration 6 2 3 Single point and multi point calibration If 1 point calibration is conducted after 3 point or 4 point calibration being done the...

Page 16: ...till be operated It is just reminding you to do calibration in order to ensure the accuracy After calibration the Er6 icon will disappear To make it disappear users can also choose No in P2 3 in param...

Page 17: ...0 100 mS cm 0 94 Sample test 1 Rinse conductivity electrode in distilled or pure water dry it and submerge it in the sample solution Stir the solution briefly and allow it to stay in the sample soluti...

Page 18: ...n for 5 10 minutes or to soak it in tap water for 1 to 2 hours Rinse the electrode in pure water after measurement 6 6 2 The sensing rod of Model 2301T S conductivity electrode is coated with platinum...

Page 19: ...Select electrode s constant P2 2 Set up calibration reminder P2 3 Check calibration date P2 4 Select reference temp P2 5 Adjust temp coefficient P2 6 Adjust TDS factor P2 7 Restore to factory default...

Page 20: ...Check Calibration Time 1 The data and time of last calibration displayed on the left image is 18 08 06 08 2016 2 Press to enter P1 4 or press to return to measurement mode P1 4 Select pH calibration...

Page 21: ...P2 5 or press to return to measurement mode P2 5 Adjust temp compensation coefficient 0 00 9 99 1 Press 2 00 flashes press or to adjust from 0 00 9 99 press to confirm 2 Press to enter P2 6 or press t...

Page 22: ...n time unit is minute 2 Press to enter P4 5 or press to return to measurement mode P4 5 Clear storage No Yes 1 Press No flashes press to choose No Yes press to confirm No not to clear data storage Yes...

Page 23: ...Port is the number for the software to connect with the computer Every computer s COM port is different In Figure 13 the COM port is COM4 Open Close Click it to turn on off the program When turned on...

Page 24: ...rage a When the program is running and meter is connected to computer press on the meter or set up auto timing storage to upload all the measuring data to the software The data will not be stored in t...

Page 25: ...rol 11 APPENDIX 1 TABLE OF PARAMETER SETTING AND FACTORY DEFAULT SETTING Mode Symbol Parameter Abbreviatio n Content Factory Default P1 0 pH P1 1 Select Buffer Series USA NIST P1 2 Set calibration rem...

Page 26: ...e P1 5 Factory default setting P2 0 Conductivity P2 1 Cell P2 2 Due Calibration P2 3 P2 4 Reference temperature P2 5 Temperature compensation coefficient P2 6 TDS Factor P2 7 Factory default setting P...

Page 27: ...value is not stable during calibration During calibration the measuring value is not stable for 3 min pH electrode zero electric potential out of range 60mV or 60mV pH electrode slope out of range 85...

Reviews: