2301-C Conductivity Electrode
User Manual
Brief Introduction
Apera Instruments 2301-C Conductivity Electrode is featured with a proprietary platinum black sensor
for accurate readings in a wide conductivity range up to 200 mS/cm.
Technical Specifications
Range
0 to 200 mS/cm
Housing
POM
Constant
K=1.0
Connector
BNC
Cable length
3 Ft
Dimension
ø12*155 mm
Temperature sensor
N/A
Operating temperature
32 to 176°F (0 - 80°C)
How to Install the Electrode
1. Find the conductivity BNC socket (where it shows Cond) on the conductivity meter; Open the
rubber cap; Insert the BNC connector of the electrode to the BNC socket while twisting clockwise
until it’s locked.
2. After connecting the new electrode to your meter,
adjust the meter’s co
nstant K to 1.0 according
to the instruction of your meter’s manual. Then perform
a 2 to 3-point calibration to ensure the
accuracy.
How to Use the Electrode
1. There is a proper amount of deionized water in the cap on top of the electrode. The conductivity
sensor is soaked in it to keep its sensitivity.
2. Before measuring, loosen the bottle cap, then pull out the electrode while twisting
counterclockwise. Place the storage bottle at a safe position.
3. Rinse the electrode with distilled or deionized water and blot-dry it with clean tissue or filter paper.
Rinse the electrode again with your test solution if possible for the best result.
4. Insert the electrode into your sample solution and stir it for a few seconds in the solution to
remove potential air bubbles. Then wait for the stable reading and take the measurement.