Paramet er Setting  

 

Table of Settings 

 

Parameter Setting 

1) 

When the meter is turned off, long press           

to enter parameter setting →short press            to switch P01-P02… 

→P8. Short Press          , parameter flashes → short press            to adjust parameter → short press            to confirm 

→Short press          to exit parameter setting and go back to measurement mode. 

2) 

Auto. Lock (P02)

 

– Users can set the auto lock time from 5 to 20 seconds. For example, if 10 seconds is set, when 

the measured value is stable for more than 10 seconds, the measured value will be automatically locked, and the 

HOLD icon will be displayed. Short press          to release the lock. When the setting is "Off" , the Auto. lock function 

is turned off, that is, the measured value can only be manually locked. Short press          to lock or unlock the 

measured value. The HOLD icon will be displayed when reading is locked. 

3) 

Auto. Backlight (P03)

 

  Users can set the automatic backlight time for 1 to 8 minutes. For example, if 3 minutes 

is set, the backlight will turn off automatically after 3 minutes; when the 

“Off” is set, the auto. backlight function will 

be turned off, and short press          to manually turn the backlight on or off. 

4) 

Auto. Power off (P04)

 

  The auto. power off time can be set to 10 to 20 minutes. For example, if 15 minutes is 

set, the meter will automatically shut down after 15 minutes if no operation; when 

“Off” is set, the auto. power off 

function will be turned off. Long press          to manually shut down the meter. 

5) 

pH Calibration Reminder (P07) 

– set X hours (H) Or X days (D) in ZenTest mobile app – settings – Parameter – 

pH 

– Calibration Reminder. On the meter, you can only check the values that’s been set up on ZenTest App. For 

example, if 3 days is set up, the Er6 icon (see Figure-4) will appear in the lower right corner of the LCD screen in 3 

days to remind you to perform calibration, also in the ZenTest App there will be a pop-up reminder. After calibration 

is finished or the reminder setting is cancelled in the ZenTest App, the Er6 icon will disappear. 

6) 

pH Back to Factory Default (P08) 

– Select “Yes” to recover instrument calibration to theoretical value. This 

function can be used when instrument does not work well in calibration or measurement. Calibrate and measure 

again after setting the instrument back to factory default.   

 

Symbol 

Parameter Setting Contents 

Content 

Factory Default 

P1 

Temperature Unit 

˚C – ˚F 

˚F 

P2 

Select automatic lock 

5-20 seconds 

– Off 

Off 

P3 

Automatic Backlight Off 

1-8 minutes

 

– Off 

P4 

Automatic Power Off 

10-20 minutes 

– Off 

10 

P5 

pH Buffer Series Selection 

USA 

– NIST 

USA 

P6 

pH Resolution 

0.1 

– 0.01 

0.01 

P7 

pH Calibration Reminder 

H-hours D-Days (set up in ZenTest App) 

P8 

pH back to factory default 

No

 

– Yes 

No 

Summary of Contents for ZenTest PH60S-Z

Page 1: ...1 ZenTest PH60S Z Smart Spear pH Tester pH Temp ORP User Manual APERA INSTRUMENTS LLC aperainst com V1 5...

Page 2: ...E 5 PH CALIBRATION 5 PH MEASUREMENT 7 ORP MEASUREMENT 8 PROBE CLEANING 8 PROBE STORAGE 8 PARAMETER SETTING 9 TECHNICAL SPECIFICATIONS 10 ICONS AND FUNCTIONS 10 TROUBLESHOOTING GUIDE 11 PROBE REPLACEME...

Page 3: ...erate the tester without connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mode digital display cal...

Page 4: ...to turn on off backlight 1 In measurement mode long press to turn on off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 3 In parameter setting short press...

Page 5: ...water RO water or tap water is the alternative then shake off excess water see graph on the right 5 3 Perform at least a 2 point calibration For pH calibration tutorial refer to Section 6 5 4 When you...

Page 6: ...librate 2nd or 3rd point Otherwise you will need to restart the calibration process with 7 00 pH first 2 The pH 4 00 and 7 00 buffer solutions poured into the calibration vials can be used for up to 1...

Page 7: ...d for over a year and has a much slower response A replacement is needed pH probe zero electric potential out of range 60mV or 60mV 1 Check whether pH buffer solutions comply with the USA or NIST stan...

Page 8: ...inal probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 15 for how to replace a probe 3 Rinse the probe in distilled water and dry...

Page 9: ...if 15 minutes is set the meter will automatically shut down after 15 minutes if no operation when Off is set the auto power off function will be turned off Long press to manually shut down the meter...

Page 10: ...unctions Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6 Stable reading indicator Waterproof Rating IP67 floats on water Reading Lock HOLD Power DC3V AAA batteries 4 Bluetooth Signal B...

Page 11: ...ble damage replace the probe Instrument defect Contact us for warranty fulfillment Probe is not fully submerged in the solution Make sure the probe is immersed into solution at least 1 inch Air bubble...

Page 12: ...ged product attributable to responsibility of APERA INSTRUMENTS LLC for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to Acciden...

Reviews: