- 12 - 

 

Probe Replacement  

pH p

robes don’t last forever. Every probe will eventually age and fail even if you don't use it that often. The typical service 

life of Apera probes is 12-24 months depending on the frequency of usage and how well you keep it clean and properly 

stored. We recommend replacing your probe every 1 to 2 years to ensure the best performance. 

 

To replace the probe: 1) take off the probe cap; 2) screw off the probe ring 3) unplug the 

probe; 4) 

plug in the new replacement probe (pay attention to the probe’s position); 5) screw 

on  the  probe  ring  tightly.  Soak  the  probe  in  3M  KCL  for  5-15  minutes.  Then  perform 

calibration before testing. 

 

The replacement probes that are compatible with PH60S-Z: 

 

PH60S-DE (Spear pH probe for solids/semi-solids pH testing), PH60-DE (Regular 

pH glass bulb probe), PH60F-DE (Flat pH probe for surface pH testing), ORP60-DA (ORP Probe). 

 

Warrant y  

We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of 

charge, at option of APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to responsibility of 

APERA INSTRUMENTS, LLC for a period of TWO YEARS (SIX MONTHS for the probe) from the delivery.   

 

This limited warranty does NOT cover any damages due to: 

Accidental  damage,  transportation,  storage,  improper  use,  failure  to  follow  the  product  instructions  or  to  perform  any 

preventive maintenance, unauthorized repair or modifications, normal wear and tear, or other external causes or actions 

beyond our reasonable control. 

 

To get the fastest warranty fulfillment, go to 

support.aperainst.com

 and click 

“New Support Ticket”

 on the upper right 

corner. Type your email in the requester field, “Warranty” in the Subject field, and then input the following information in 

the description field: 

 

Your full name 

 

Product model that needs warranty fulfillment 

 

Serial number of the product (can be found on the back sticker of the tester body) 

 

What problem or issue you had experienced with the product 

 

Attach a photo of your proof of purchase 

 

Attach a photo of the problematic product 

Then click Submit. One of our customer care specialists will help you fulfill the warranty within one business day. 

 
 
 
 
 

APERA INSTRUMENTS, LLC 

Address: 6656 Busch Blvd, Columbus Ohio 43229

Tel: 1-614-285-3080 

Email

[email protected]

   

Website: www.aperainst.com 

Summary of Contents for ZenTest PH60S-Z

Page 1: ...1 ZenTest PH60S Z Smart Spear pH Tester pH Temp ORP User Manual APERA INSTRUMENTS LLC aperainst com V1 5...

Page 2: ...E 5 PH CALIBRATION 5 PH MEASUREMENT 7 ORP MEASUREMENT 8 PROBE CLEANING 8 PROBE STORAGE 8 PARAMETER SETTING 9 TECHNICAL SPECIFICATIONS 10 ICONS AND FUNCTIONS 10 TROUBLESHOOTING GUIDE 11 PROBE REPLACEME...

Page 3: ...erate the tester without connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mode digital display cal...

Page 4: ...to turn on off backlight 1 In measurement mode long press to turn on off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 3 In parameter setting short press...

Page 5: ...water RO water or tap water is the alternative then shake off excess water see graph on the right 5 3 Perform at least a 2 point calibration For pH calibration tutorial refer to Section 6 5 4 When you...

Page 6: ...librate 2nd or 3rd point Otherwise you will need to restart the calibration process with 7 00 pH first 2 The pH 4 00 and 7 00 buffer solutions poured into the calibration vials can be used for up to 1...

Page 7: ...d for over a year and has a much slower response A replacement is needed pH probe zero electric potential out of range 60mV or 60mV 1 Check whether pH buffer solutions comply with the USA or NIST stan...

Page 8: ...inal probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 15 for how to replace a probe 3 Rinse the probe in distilled water and dry...

Page 9: ...if 15 minutes is set the meter will automatically shut down after 15 minutes if no operation when Off is set the auto power off function will be turned off Long press to manually shut down the meter...

Page 10: ...unctions Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6 Stable reading indicator Waterproof Rating IP67 floats on water Reading Lock HOLD Power DC3V AAA batteries 4 Bluetooth Signal B...

Page 11: ...ble damage replace the probe Instrument defect Contact us for warranty fulfillment Probe is not fully submerged in the solution Make sure the probe is immersed into solution at least 1 inch Air bubble...

Page 12: ...ged product attributable to responsibility of APERA INSTRUMENTS LLC for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to Acciden...